Calibration and Services – Polarimeter Calibration

ZERO Calibration
Polarimeters are generally zero calibrated using air or distilled water. However, in some cases it is desirable to use an active blank solvent to zero the instrument, allowing further measurement to reflect any optically active component that is later added to the solvent.

Bellingham + Stanley Ltd. do not supply any materials for zero calibration.

SPAN Calibration - Sucrose Solutions
Sucrose Solutions of known concentration may be used to verify a polarimeter operated in either the angular or sugar scale. Sucrose solutions are primarily used in the sugar and food industries but are considered too time consuming to prepare, requiring a new sample to be made for each calibration as its optical rotation is affected by time. Sugar Solutions are also temperature dependant so it is desirable to use a temperature controlled polarimeter tube or employ temperature compensation if available.

SPAN Calibration - Quartz Control Plates
Bellingham + Stanley Ltd. offer a choice of Quartz Control Plates (QCP) for verifying and calibrating polarimeters. QCP's are made to the highest standard and may be supplied with an optional Certificate of Calibration, showing traceability to PTB. When used with an ADP polarimeter manufactured by Bellingham + Stanley, a thermal block may be used to provide mechanical contact to the instrument's external temperature sensor, allowing the use of quartz temperature compensation for added accuracy.

Order Code Nominal Value Description
34-20 +100 °Z +34.6 °A Standard Quartz Control Plate supplied in soft protective cover, Certificate of Conformity and packed in a rigid case for transit and storage.
34-21 +15 °Z +5.2 °A
34-22 -30 °Z -10.4°A
34-241   Thermal Block for use with ADP/S
Uncertainty (new plates): ±0.019 °Z and ±0.007 °A

Certification Service
UKAS Calibration number 0834Quartz Control Plates may be supplied from new with a traceable certificate, as an optional service. Additionally, as part of GLP, it is often required to have existing Quartz Control Plates recalibrated at regular intervals.

Bellingham + Stanley Ltd., a UKAS accredited Calibration Laboratory number 0834, offer a service whereby plates within the following ranges may be issued with a Certificate of Calibration, traceable to PTB:


Range ISS Angular
1 13 to 17 °Z 4.5 to 5.9 °A
2 95 to 105 °Z 32.9 to 36.4 °A
3 -26 to -34 °Z -9.0 to -11.8 °A

Uncertainty: ±0.019 °Z and ±0.007 °A



Order Code Description
90-803 Calibration, quartz control plate at 20.0 °C & 589.44nm.

Best Measurement Capability (Uncertainty):
The reported expanded uncertainty of measurement is stated as the standard uncertainty of measurement in units of °Z multiplied by the coverage factor k = 2, which for a normal distribution corresponds to a coverage probability of approximately 95%. The standard uncertainty of measurement has been determined in accordance with publication EA-4/02.

FAQ
Technical Bulletion TB-003 – Calibration of Polarimeters using sucrose.



Information Request

Dear Bellingham + Stanley

I have visited your website and am interested in receiving further information about your products and services.

Please can you therefore provide:

  • Further technical information
  • A quotation
  • Local distributor details
Company:* 
Address:*
Country:*
Postal/ZIP Code:*
 
Contact Name:*
Contact Phone Number:
Contact E-mail address:*
Product(s) of interest:*  
Message:
* indicates required information
Close Form